AEQ na quantificação da hemoglobina: desempenho analítico a longo prazo de trinta laboratórios portugueses

dc.contributor.authorMiranda, Armandina
dc.contributor.authorCorreia, Helena
dc.contributor.authorCardoso, Ana
dc.contributor.authorBrito, Cristina
dc.contributor.authorClemente, Vera
dc.contributor.authorSilva, Susana
dc.contributor.authorMeijer, Piet
dc.contributor.authorFaria, Ana
dc.date.accessioned2016-12-06T15:54:48Z
dc.date.available2016-12-06T15:54:48Z
dc.date.issued2016-11
dc.description.abstractÉ importante para os laboratórios clínicos avaliarem o desempenho dos seus métodos ao longo do tempo. O principal objetivo deste estudo foi a aplicação de um modelo de regressão linear para avaliação a longo prazo, do desempenho dos laboratórios na quantificação da hemoglobina, comparando os resultados individuais do laboratório com a média de consenso de cada ensaio, após a exclusão de outliers. Para avaliar o desempenho do laboratório foram calculados o coeficiente de variação analítico a longo prazo (LCVa) e o Bias total analítico. O modelo de regressão linear foi aplicado aos resultados quantitativos da hemoglobina para avaliar o desempenho analítico a longo prazo de trinta laboratórios que participam no programa de avaliação externa da qualidade PNAEQ (2014-2016), utilizando os resultados de doze amostras controlo de sangue em EDTA, com diferentes concentrações de hemoglobina. Os participantes foram selecionados aleatoriamente, sendo incluídos laboratórios hospitalares e ambulatórios. As variáveis introduzidas para avaliar o desempenho a longo prazo neste modelo foram o CV analítico a longo prazo (LCVa) e o Bias total analítico. Dois laboratórios foram excluídos da análise. Foi determinado o número de laboratórios que cumpriram os objetivos de desempenho analítico definido com base na variação biológica. A mediana do LCVa foi de 1,4% (variação de 0,4% -3,4%). O LCVa foi 0,58 vezes inferior à variação biológica total (teste de diagnóstico) para todos os laboratórios (100%) e foi 0,5 vezes inferior à variação biológica intraindividual (teste de monitorização) em 62% dos laboratórios. A mediana do Bias total foi de 1,1% (variação de 0,1% -3,3%), sendo que 83% dos laboratórios apresentaram um Bias total inferior a 0,25 da variação biológica total. A participação em programas de avaliação externa da qualidade é de extrema importância, uma vez que fornece informações ao laboratório sobre o desempenho do seu método tanto numa única análise, bem como ao longo do tempo e permite a avaliação da necessidade de melhoria.pt_PT
dc.description.abstractIt is impor tant for medical laboratories to know about the stability and performance of their methods over the time. The main objective of this study was to apply a linear regression model for long-term evaluation of the performance of laboratories in haemoglobin quantification by comparing the laborator y individual results with the consensus mean of each round, af ter outliers exclusion. To evaluate the per formance of the laborator y, the long term analytical CV (LCVa) and the total analytical bias were established. A linear regression model was applied to quantitative haemoglobin results to evaluate the long-term analytical per formance of thir ty laboratories that par ticipate in the PNAEQ external quality assessment programme (2014- 2016), using the results of twelve blood EDTA samples with dif ferent hemoglobin concentrations. Par ticipants were randomly selected concerning laboratories from hospital and ambulator y. The variables introduced to define the long-term per formance in this model were the long-term analytical CV (LCVa) and total analytical Bias. Two laboratories were excluded from the analysis (one outlier and one that uses point-of-care equipment). We evaluate also the number of laboratories that fulfill the widely accepted analytical per formance goals based on the biological variation. The median LCVa was 1.4% (range 0.4%-3.4%). The LCVa was less than 0.58 times the total biological variation (diagnostic testing) for all laboratories (100%) and was less than 0.5 times the within biological variation (monitoring testing) in 62 % of the laboratories. The median Total Bias was 1.1% (range 0.1%- 3.3%). 83% of the laboratories had a total bias less than 0.25 of the total biological variation. Par ticipation in External Quality Assessment schemes is of extreme impor tance, as it provides information to the laborator y about its own method per formance both in a single sur vey as well as over time and allows the evaluation of the needs for improvements.pt_PT
dc.identifier.citationBoletim Epidemiológico Observações. 2016 setembro-dezembro;5(17):29-31pt_PT
dc.identifier.issn0874-2928
dc.identifier.issn2182-8873 (em linha)
dc.identifier.urihttp://hdl.handle.net/10400.18/4109
dc.language.isoporpt_PT
dc.publisherInstituto Nacional de Saúde Doutor Ricardo Jorge, IPpt_PT
dc.rights.urihttp://creativecommons.org/licenses/by-nc/4.0/pt_PT
dc.subjectAvaliação Externa da Qualidadept_PT
dc.subjectHemoglobinapt_PT
dc.subjectLaboratórios Portuguesespt_PT
dc.subjectPrograma Nacional de Avaliação Externa da Qualidadept_PT
dc.subjectPNAEQpt_PT
dc.subjectSaúde Públicapt_PT
dc.subjectPortugalpt_PT
dc.titleAEQ na quantificação da hemoglobina: desempenho analítico a longo prazo de trinta laboratórios portuguesespt_PT
dc.title.alternativeEQA total haemoglobine quantification: long-term analytical performance for thirty Portuguese laboratoriespt_PT
dc.typejournal article
dspace.entity.typePublication
oaire.citation.conferencePlaceLisboa, Portugalpt_PT
oaire.citation.endPage31pt_PT
oaire.citation.startPage29pt_PT
oaire.citation.volume5(17)pt_PT
rcaap.rightsopenAccesspt_PT
rcaap.typearticlept_PT

Files

Original bundle
Now showing 1 - 1 of 1
Loading...
Thumbnail Image
Name:
Boletim_Epidemiologico_Observacoes_N17_2016_artigo7.pdf
Size:
668.1 KB
Format:
Adobe Portable Document Format
License bundle
Now showing 1 - 1 of 1
No Thumbnail Available
Name:
license.txt
Size:
1.71 KB
Format:
Item-specific license agreed upon to submission
Description: